Cerabar M PMP55 - Digital Pressure Transmitter With Diaphragm Seal

 

 

 

Digital pressure transmitter with diaphragm seal for measurement
in gases, steams or liquids

 

The Cerabar PMP55 digital pressure transmitter with metal diaphragm seal is
typically used in the process and hygiene industry for pressure, level, volume
or mass measurement in liquids. 
PMP55 is designed for high pressure and high
temperature applications up to 400°C and 400bar. The Quick Setup with freely
adjustable measuring range without pressure specification for simple commissioning
reduces costs and saves time. SIL2 according to IEC 61508 and IEC 61511.

 

 

Benefits

  • Large variety of different membrane materials and process connections

  • Very good reproducibility and long-term stability

  • For process pressure monitoring up to SIL2, certified to IEC 61508 and IEC 61511

  • Easy menu-guided commissioning via on-site display, 4 to 20mA with HART, PROFIBUS PA, FOUNDATION Fieldbus

  • Modular concept for differential pressure, hydrostatics and pressure transmitters, e.g. replaceable display, universal electronics

  • Seamless and independent system integration (HART/PA/FF)

  • New TempC diaphragm for diaphragm seals: Minimum temperature effects and short recovery times

 

 

 

Field of application

Single chamber aluminum or stainless steel (316L) housing.

  • Process connections: threads, flanges or hygienic connections

  • Temperature: -70 to +400°C (-94 to +752°F)

  • Pressure: +400mbar to +400bar (+15 to +6,000psi)

  • Accuracy: ±0.15%, "Platinum" ±0.075%

  • International explosion protection certificates, EN10204-3.1, marine approval, 3-A, EHEDG, CoC ASME-BPE

 

 

Technical information

Measuring Principle Hydrostatic
Characteristic Smart and reliable pressure transmitter, with piezoresistive measuring cell and metal welded process isolating diaphragm and diaphragm seal. 
Supply / Communication 4 ..20mA HART:
11,5...45V DC
Ex ia: 11,5...30V DC
PROFIBUS PA/ FOUNDATION Fieldbus:
9..32 V DC (Non Ex)
 
Accuracy Standard 0,15%
Platinum 0,075%
Long term stability 0,1% of URL/year 
Process temperature -70°C…400°C 
(-94°F…752°F) 
Ambient temperature -40°C...85°C 
(-40°F...185°F) 
Measuring range 400mbar...400bar
(15psi...6000psi) 
Max. overpressure limit 600bar (9000psi) 
Process connection Thread:
G1/2...G2, MNPT1/2...MNPT2
Flange:
DN25...DN100,
ASME 1"...4",
JIS 10k
Diaphragm seal 
Hygienic Tri-Clamp
DIN11851
DIN11864-1
NEUMO
Varivent
SMS
DRD
Universal adapter
Output 4...20 mA
4...20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus 
Certificates / Approvals ATEX, FM, CSA, CSA C/US, IEC Ex, TIIS, INMETRO, NEPSI
SIL
EN10204-3.1
NACE
CoC ASME-BPE
3A, EHEDG
Marine
Options Display
PWIS free 
Specialities TempC Membrane
  •